Victoreen 875 High Range Containment Monitor
Operator Manual

2-4

Ancillary Wiring

Ancillary wiring for computer, recorder, power and alarm contacts are installed accor
875-1 and Table 2-1 which indicates connector pin designations.

able 2-1. Installation

Computer & Recorder

ding to drawing GEL

T

Function

J3

Ground

D

Ground

F

Computer

E

Recorder

C

All Alarms on P2

Normal Operation *

Shelf State

A

A

lert larm I

N

NO

NC

P

C

C

R

NC

NO

Alert A

NO

NC

larm II

D

E

C

C

F

NC

NO

High Alarm I

S

NO

NC

T

C

C

U

NC

NO

High Alarm II

G

NO

NC

H

C

C

J

NC

NO

Fail Alarm I

K

NO

NC

L

C

C

M

NC

NO

Fail Alarm II

A

NO

NC

B

C

C

C

NC

NO

Power Connections (P1)

A

Line

Neutral

B

Chassis

Ground

C


* Rela

NO =

C

N

ys are Fail Safe, i.e. energized for normal operation. Relays de-energize during an alarm condition.

Normally open

= Common

C = Normally closed