3 troubleshoo, Ting – Fluke Biomedical 875 Victoreen User Manual

Page 34

Advertising
background image

Maintenance, Calibration, and Troubleshooting

Troubleshooting

5


5-3

Test Procedure TP876A-1-108

Factory test procedure,

rm electronic

alignment of the 876A-1

dix A.

5.3 Troubleshoo

The 875 High-Range Co

r has been assembled

y techniques and with parts selected for the reliability required in a nuclear application. Any repairs

safety-

m.

ircuit boards must be returned to

There are two self-contained system tests avai

inment Monitor 875, the Channel Test and

the ECS Test. In both ca
desired output.

mportant di

Test and the ECS Test: the former applies an input

the readout module (that is, to the first electronic circuit in the Containment Monitor); however, it does

plates,

libration section is also recommended for

If

4.

lfunction exists in the fail/safe circuitry.

Essentially, the Ch

ntally verifies all

o

1.

ircuit

4. Alarm

circuits

6.

TP876A-1-108, contains this information necessary to perfo

-108 Readout Module. This procedure is included in Appen

ting

ntainment Monitor is safety-related equipment. The monito

b
made to the detector or readout (other than replacement of parts listed in Section 5) may void the
related rating. The troubleshooting procedure that follows is a guide to isolating a fault in the syste
Replacement of parts is at the printed circuit board level only. Printed c
Fluke Biomedical for service.

lable in Conta

ses, the procedure is to put a known input into the system, and to look for the

fference between the Channel

There is an i
to
not test the detector or the cables connecting the detector to the readout module. In addition, the Channel
Test applies a DC voltage, whereas the ECS Test applies a ramp voltage to the detector
monitoring the resultant current into the readout module.

The digital multimeter mentioned in Table 5-1 of the ca
troubleshooting.

the ECS Test gives a favorable result, the following conditions exist:

1. The detector cannot have any appreciable malfunction.

2. The cables must have continuity.

3. The ECS board and power supply board must be operative.

It is highly improbable that any ma

5. The amplifier, meter and alarm circuitry must be operative.

annel Test verifies only #5 of the above. However, in doing so, it incide

f the power supplies except the high voltage power supply.

For clarity, troubleshooting is divided into six divisions:

Power supplies

2. Input

c

3. Meter

circuit

5. ECS

board

Overall fail circuitry

Advertising