Victoreen 875 High Range Containment Monitor
Operator Manual

5-12

able 5-1. Truth Table for Pass and Fail Conditions in The Latch Circuit

(Circuit assumed initially in Pass Condition)

T

Fail Condition

Pass Condition

Pin

Voltage
State

Voltage Comments

Pin

State

Comments

4

L (low)

Since 6 is H (high)

4

H

Because 6 is L (also 5)

3

H (high)

Since 1 & 2 are Low

3

L

Because 2 is H

13

H

Connected to 3

8

H

Connected to 4

11

L

Since 13 is H

13

L

Connected to 3

9

L

Connected to 11

10

L

Because 8 is H

8

L

12

L

Connected to 4

10 H

9, 10) is

11

H

Bec

Since U202, (8,
NOR

ause 12 and 13 are L

12 H

from Lo

Result: Output switches

w to High

This is the original condition. No
change takes place.


Notes:

ition assumes 12 and 1

The truth table shows that no change in the output can occur without the presence of the enable pulse,

the monitoring signal indicat

the ramp voltage

able 5-2. Signal In Fail State (High Signal Input)

Pin 1

Voltage State

Comments

1) Enable pulse assumed present, so 1 and 5 are L.

2) Initial cond

0 L.

even though

es a fail condition. This is necessary because

starts from zero during the ECS Test.

T

10, U208 and 6, U202

H

Fail state

9, U208 and 2, U202

L

After inversion

1,5, U202

H

No enable input

4, 8, U202

L

Because of NOR action

3, 13, U202

L

Because of NOR action

10, 12, U202

L

Because of NOR action


Thus the circuit does not react to a fail signal input unless pins 1 and 5 are low.



If a problem cannot be resolved by applying the
troubleshooting and maintenance procedures
provided here, please contact Fluke Biomedical at
440.248.9300 for assistance.