Section 5, Maintenance, calibration, and troubleshooting, 1 maintenance – Fluke Biomedical 875 Victoreen User Manual

Page 32: Calibration 5.2 field

Advertising
background image

Maintenance, Calibration, and Troubleshooting

Maintenance

5

5-1

Section 5

Maintenance, Calibration, and Troubleshooting

r

n the containment without attention. The following

e

uble-free operation.

very five years - Replace the 4

capacitor, P/N 92-3005-A (C1 01), and the RFI

e filter, P/N 92-9015A (FL101

other board of the 876A-1-108 or 876A-100

henever detector cables are re

re in Appendix A.

e

libration before placing them into service. In addition, they should be

als

ring routine service. The length of time between calibration intervals

determined by operat

r to the applicable

procedure provided in

he high-range containment are

calibration prior

e Biomedical

c calibration procedure for the 876A-1 Readout is

is manual. Prior to

topic calibration, the detector's hermetic seal, insulation

d capacitance are

opic calibration requires a highly radioactive source

400 curies) with National Bureau of Standards (NBS) traceability. As this is beyond the

ost facilities to perform, the following method of verifying detector calibration is

recommended:

d to Fluke Biomedical at a five (5) year interval from the date of delivery

er must establish a

that the average A/R/h output current or field

se, does not

factory calibration by more than

± 20 %.

reg-0737 guidelines, on-site in-situ calibration checks can be performed with the 878-

ld Calibrato

producing a 10 R/h indication on the channel under

ducibility of the

in +/- 20 %.

878-10 Field C

he 877-1 Detector, its reproducibility is based on the

aracteristics o

e, the mechanical tolerances of the calibrator, detector

ll strin

skill of the user. Factory recalibration is recommended

e decayed output value exceeds +/- 20 %, or if the unit has been physically

alignment of the 87

nal Test Procedure,

-108 included in App

-1.

nnel must be fully knowl

pting to perform the

test. The Readout should be c

rce to verify its ability to measure the input

rately to within +/-

5.1 Maintenance

The monitor is designed to ope
replacement schedule should k

ate for long periods i

ep the monitor in tro

E

200 microfarad electrolytic

lin
R

), located on the 876-1-78 m

eadout Module.

W
co

moved - Replace the nickel seals, P/N 877-1-60-1, on the detector

nnectors as described in the CABLE-877 procedu

Calibration

5.2 Field

The monitoring systems req
recalibrated at regular interv

uir ca

du

should be

ions personnel. For further calibration information, refe

calibration

Appendix A.

T

a monitor underwent a complete electronic and isotopic

to leaving the Fluk

plant. The electroni

supplied in th

primary iso

. Primary isot

resistance an
(greater than

verified

capability of m

Detectors shall either be returne
or the own

procedure to determine

calibrator respon

deviate from original

To encompass Nu

Fie

10 High-Range
test. The repro

r that is capable of

calibrator is with

Concerning the
known decay ch

alibrator used with t

f the

137

Cs sourc

positioning, calibrator pu

g tension and the

if the reproducibility of th
damaged.

Electronic
TP876A-1

6A Readout may be performed using Functio

endix A, and the standard test equipment as listed in Table 5

edgeable in the operation of the readout prior to attem

Perso

alibrated with a current sou

current accu

10 %.

Advertising