Victoreen 875 High Range Containment Monitor
Operator Manual

5-2

NOTE

Detector 877-1 from Readout Module

t Equipment

Disconnect
876A-1 by removing cables from J1 and J2 before
starting electronic calibration.

Table 5-1. Recommended Tes

Current Generator - Test Electrometer

Range

1 x 10

-12

Amps to 1 x 10

-3

Amps

Accuracy, Current Generator

10

-12

Amps

± 10 %

10

-11

Amps

± 3 %

-10

-7

10

to 10 Amps

± 2 %

10

-7

to 10

-3

Amps

± 0.2 %

Accuracy Test Electrometer

± 0.2 % ± 1 digit with integration time long enough to accumulate 3
significant digits of volts indicated on display V/R

± 0.2 % ± 1 digit

Adjustable AC Voltage Source

Maximum Voltage

Low range 140 VAC, high-range 280 VAC

Frequency 60

Hz

4 ½ Digit Digital Multimeter

Fluke 8600A or equivalent

Ranges

+ 0 to 100 mV up to + 0 to 100 V (20 % over-range)

Accuracy

100 mV range:

± (0.05 % of input ± 0.02 % of range)

all other ranges:

± (0.02 % of input ± 0.01 % of range)

In

put Impedance

> 10 megohms

D

C Voltage

± 0.03 %

AC Voltage

± 1 %

DC Current

± 0.3 %

Trim Pot Adjustment Tool

Carbon Composition Resistor

Resistance

10 kilohms, ½ W, 5 %

Lead Connectors

Small alligator clips

DC Power Supply

Output Voltage

0 to + 25 VDC

Output Current

0 to 600 mA

L

0.01 %

± 4 mV

oad Regulation

A

it

Down to 10 mA

djustable Current Lim

M

nter

ulti-function Cou

F

luke 1900A or equivalent

Frequency

5 Hz to 80 MHz (0.1 to 100 Hz Resolution)

Period

5 Hz to 1 MHz, single and multiple period averages

(1 to 100 nsec resolution)

Total Counts

1 to 999,999 counts