Pacemaker selections – Fluke Biomedical PS410 User Manual

Page 13

Advertising
background image

ECG Simulator

Specifications

5

Standard Accessories ..........................................Users Manual (printed) (PN 2631795)

Users Manual CD-ROM (PN 2631742)
Battery Eliminator (PN 2647372)

ECG

12 Leads with Independent Outputs Referenced to RL

Output Impedance...............................................940 ohms between leads
High Level Output................................................1000x Lead II
Rates ...................................................................30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, and 300 BPM
Default Rate ........................................................80 BPM
Rate Accuracy .....................................................±1 % of selection

Adult or Pediatric Waveforms

ECG Amplitudes..................................................0.5, 1.0, 1.5, and 2.0 mV
Amplitude Accuracy.............................................±2 % (Lead II).
Superimposed Artifact .........................................50 and 60 Hz, muscle, baseline wander, and respiration.

ECG Performance, Lead II

Square Wave.......................................................0.125 and 2.0 Hz.
Pulse ...................................................................30, 60, and 120 BPM; 60 ms pulse width.
Sine Waves .........................................................0.5, 5, 10, 40, 50, and 60 Hz (1 mV amplitude).
Triangle Wave .....................................................2.0 Hz.

ST Segment Analysis

Elevated or Depressed........................................-0.2 mV to +0.6 mV in 0.2 mV steps

Pacemaker Selections

Pacemaker Rhythm

Demand Occasional Sinus

Pacemaker Non-Function

Demand Frequent Sinus

Pacemaker Non-Capture

A-V Sequential