Superimposed artifacts, Pacemaker – Fluke Biomedical PS410 User Manual

Page 24

Advertising
background image

PS410
Users Manual

16

Superimposed Artifacts

The Simulator replicates five different artifacts. Their
purpose is to evaluate the effect of these artifacts on ECG
accuracy. After selecting the artifact, press ENTER to
transmit it to the ECG. The Simulator deactivates the
artifact when you make another ECG or arrhythmia
selection.

Code Display

Selects:

75

50 Hz AR

50 Hz artifact (European lines).

76

60 Hz AR

60 Hz artifact (US lines).

77 MUSCLE Muscle

artifact.

78

BASE ART

Baseline wandering artifact.

79

RESP ART

Respiration artifact.

Pacemaker

The Simulator replicates six paced rhythms/signals. After
selecting the required rhythm, press ENTER. After
selecting the preset, press ENTER. To change the rhythm,
use the UNITS keys ( ) to scroll to the available presets.

Code

Display Selects:

63

PACER RH

Pacemaker rhythm.

64

PACER NC * Noncapture event. The Simulator

then assumes asynchronous
pacemaker.

65

PACER NF

Non-function pacemaker rhythm.

66

DEMAND S

Pacemaker rhythm with
occasional sinus beats.

67

DEMAND F

Pacemaker rhythm with frequent
sinus beats.

68 SEQUENTA

Atrial-Ventricular

sequential

pacemaker rhythm.

* The simulation event occurs only once. To repeat the
event, enter the selection again.