ECG Simulator

Simulating Functions

17

ECG Performance Testing

Square Wave

Code Display

Sets:

22

2 Hz

2.0 Hz square waveform.

23

0.125 Hz

0.125 Hz square waveform.

Triangle Wave

Code Display

Sets:

24 TRI 2 Hz

2.0 Hz triangle waveform.

Pulse Wave

Code Display

Sets:

25

30 BPM P

Pulse of 30 BPM, width of 60 ms.

26

60 BPM P

Pulse of 60 BPM, width of 60 ms.

27

120BPM P

Pulse of 120 BPM, width of 60 ms.

Sine Wave

The Simulator fixes amplitude at 1.0 mV for sine waves.

Code

Display Selects:

69

0.5 Hz SI

0.5 Hz sine wave.

70

5 Hz SI

5.0 Hz sine wave.

71

10 Hz SI

10.0 Hz sine wave.

72

40 Hz SI

40.0 Hz sine wave.

73

50 Hz SI

50.0 Hz sine wave.

74

60 Hz SI

60.0 Hz sine wave.