Ecg performance testing, Square wave, Triangle wave – Fluke Biomedical PS410 User Manual
Page 25: Pulse wave, Sine wave
Advertising

ECG Simulator
Simulating Functions
17
ECG Performance Testing
Square Wave
Code Display
Sets:
22
2 Hz
2.0 Hz square waveform.
23
0.125 Hz
0.125 Hz square waveform.
Triangle Wave
Code Display
Sets:
24 TRI 2 Hz
2.0 Hz triangle waveform.
Pulse Wave
Code Display
Sets:
25
30 BPM P
Pulse of 30 BPM, width of 60 ms.
26
60 BPM P
Pulse of 60 BPM, width of 60 ms.
27
120BPM P
Pulse of 120 BPM, width of 60 ms.
Sine Wave
The Simulator fixes amplitude at 1.0 mV for sine waves.
Code
Display Selects:
69
0.5 Hz SI
0.5 Hz sine wave.
70
5 Hz SI
5.0 Hz sine wave.
71
10 Hz SI
10.0 Hz sine wave.
72
40 Hz SI
40.0 Hz sine wave.
73
50 Hz SI
50.0 Hz sine wave.
74
60 Hz SI
60.0 Hz sine wave.