ECG Simulator

Simulating Functions

15

Arrhythmias: Conduction

Defects

Code Display

Selects:

58

1ST DEGR 1st degree heart block rhythm.

59

2ND DEGR 2nd degree heart block rhythm.

60

3RD DEGR 3rd degree heart block rhythm.

61

RBBB

Right bundle branch block rhythm.

62

LBBB

Left bundle branch block rhythm.

ST Elevation and Depression Waves

These are set through direct code entry, as follows. After
selecting the preset, press ENTER. To change the wave,
use the UNITS keys ( ) to scroll to the available presets.

Code

Display Selects

ST elevation wave of:

80 ST+.6

mV

+0.6 mV

81

ST+.4 mV

+0.4 mV

82

ST+.2 mV

+0.2 mV

Selects ST depression wave of:

83

ST-.2 mV

- 0.2 mV

84

ST-.4 mV

- 0.4 mV

85

ST-.6 mV

- 0.6 mV