Code Display

Selects:

33

RUN 11 *

Run of 11 PVCs. The Simulator
then assumes NSR at 80 BPM.

35

FREQ MUL

Frequent multifocal rhythm.

39

BIGEMINY

Bigeminy rhythm.

40

TRIGEMIN

Trigeminy rhythm.

41

PVC 6/M

6 PVCs per minute.

42

PVC 12/M

12 PVCs per minute.

43

PVC 24/M

24 PVCs per minute.

44

VENT TAC

Ventricular tachycardia.

45

VENT FIB

Ventricular fibrillation.

46

V FIB #2

Ventricular fibrillation at 1/2.

47

SUPRA VE

Supraventricular tachycardia.

50

ASYSTOLE

Asystole. No ECG.

* The simulation event occurs only once. To repeat the
event, enter the selection again.

Arrhythmias: Atrial

Code

Display Selects:

48

PAC ATRI *

Premature atrial contraction. The
Simulator then assumes NSR at
80 BPM.

49

PNC NODA *

Premature nodal contraction.
The Simulator then assumes
NSR at 80 BPM.

51

MISSED B *

Missed beat. The Simulator then
assumes NSR at 80 BPM.

52

NODAL RY

Nodal rhythm.

53

IRREG RY

NSR with irregular rhythm.

54

A FLUTTE

Atrial flutter.

55

A FIBRIL

Atrial fibrillation.

56

A FIB #2

Atrial fibrillation 1/2 value.

57

A TACHYC

Atrial tachycardia.

* The simulation event occurs only once. To repeat the
event, enter the selection again.