Getting started, Power up – Fluke Biomedical 05-450 User Manual

Page 15

Advertising
background image

Digital Doorway Monitor

Getting Started

7

Getting Started

The Product is designed for ease of use. This section of the manual is designed

to help the first-time user get started. Initial power-up and basic features of the

instrument are contained in this section. Other sections of the manual provide

more detailed information

Power Up

Plug the power cord into a suitable wall (Mains) outlet.

Note

The Product will normally be wired internally for 120 V ac. If
requested, the Product may be wired for 220 V ac Check the label
next to the ac input receptacle to verify the required input voltage.

If the RS-232 feature is used, plug in a suitably wired 9-pin connector cable. See

the section 9-Pin Data Connector. for pin assignment. Turn power ON with the

left side panel switch. Do not turn power OFF unless the unit is to be removed

from service.
Figure shows the left side-panel.

INPUT:120 VAC 50/60 HZ 250W

Use only with a 250V fuse

LINE FUSE: 2 EACH LITTEL FUSE F1A, L250V

Figure 2. Left Side-Panel

Read and then remove the sticker (see Figure ) from the instrument calibration

cover. Checking and setting of the alarm point(s) is discussed in detail below and

in the sections Setting Alarm-Points and Calibration Controls in this manual.

Advertising
This manual is related to the following products: