Calibration, High voltage, Calibration parameters – Fluke Biomedical 05-450 User Manual

Page 23

Advertising
background image

Digital Doorway Monitor

Calibration

15

Calibration

High Voltage

The high voltage is adjustable from 600 V dc to 1200 V dc using the HV

potentiometer located under the calibration cover. Ensure that the high voltage is

checked only with a high impedance (≥1000 megohm) voltmeter only. A high-

voltage checkpoint is located next to the HV potentiometer. The high voltage

required will depend on the type of detectors used. Normally, each detector in a

two detector system is separately plateaued and documented. The two plateau

sheets are then compared and an operating voltage is selected that is compatible

to both detectors.

Warning

To prevent possible electrical shock, fire, or personal injury do
not touch the circuit board in the calibration window due to
potential for electric shock.

Calibration Parameters

The calibration parameters, LOW ALARM, HIGH ALARM, CAL CONST, and

DEAD TIME can only be changed while in calibration mode. Switch the top

dipswitch CAL MODE to the right to switch into calibration mode. Changing any

parameter is done by holding down the parameter key and pressing the up or

down arrow buttons. Any parameter can be set in the range of 0.1 to 9999. If a

parameter is changed, the instrument will beep to confirm the saving of the

parameter, and then return to displaying the current radiation level.
The calibration constant (CAL CONST) is set when the detector is exposed to a

"low" radiation field. A "low" radiation field in this case is defined as a field where

dead time losses do not exceed 5 %. The calibration constant is usually given for

a certain detector. Once the calibration constant is set and checked at a low

radiation field, the dead time correction can be set.
The dead time correction (DEAD TIME) is set when the detectors are exposed to

a "high" radiation field. A "high" radiation field in this case is defined as a field

where dead time losses exceed 30 %. The dead time correction will elevate the

ratemeter reading to account for counts arriving at the detector during the

detector's dead time. Scintillation detectors generally have short dead times from

1 microsecond to 10 microseconds.

Note

Once parameters are set, it is important to remember to switch the
CAL MODE switch back to the left. This action protects the
parameters from inadvertent changes.

Advertising
This manual is related to the following products: