Dipswitch (under calibration cover) – Fluke Biomedical 05-450 User Manual

Page 19

Advertising
background image

Digital Doorway Monitor

Operator Controls and Setup

11

Dipswitch (under calibration cover)

When the calibration cover is removed, a four-pole dipswitch is accessible that

can activate or deactivate options. These four options are: CAL MODE, LATCH
ALARM

, RANGE, and SINGLE BEEP.

Dipswitch 1: Switching the top CAL MODE switch to the right places the

instrument into calibration mode. Parameters can only be changed while the

instrument is in calibration mode. Calibration mode also changes the analog

output to full-scale so that the full-scale voltage may be set by the ANALOG

potentiometer. Calibration mode also slows the response time of the display

and increases the accuracy. If the display seems too erratic, leaving this

switch in the calibration mode during operation will help. Moving the CAL
MODE

switch back to the left locks the parameters and disables any further

changes.

Dipswitch 2: The second switch, LATCH ALARM, changes the high alarm to

a latching alarm. This switch does not affect the low alarm, which is always

non-latching. When switched to the left, the high alarm is non-latching; the

alarm automatically turns off when the radiation level drops below the alarm

point. When switched to the right, the high alarm light and audio signals are

latched until either the LOW ALARM or HIGH ALARM button is pressed.

Dipswitch 3: The third switch, RANGE, selects the range of the instrument.

To select the 0.1 <units> -999.9 <units> range, switch the

RANGE

switch to

the left. To select the 1 <units> -9999 <units> range, switch the RANGE

switch to the right.

Dipswitch 4: Switching the fourth switch to the right places the instrument

into SINGLE-BEEP mode. This option limits the audio output to a single half-

second beep on LOW ALARM and HIGH ALARM. DET FAIL audio output

(steady tone) is not limited.

Advertising
This manual is related to the following products: