Pin data connector, Detector setups – Fluke Biomedical 05-450 User Manual

Page 21

Advertising
background image

Digital Doorway Monitor

Operator Controls and Setup

13

9-Pin Data Connector

The 9-pin connector provides output signals from the instrument and input

voltage to the instrument. The pin assignments are:
pin1-

+BATTERY

pin2-

GND

IN

pin3-

FAIL_L

pin4-

RS232

DUMP

pin5-

ANALOG

OUT

pin6-

NA

pin7-

HIGH

ALARM_L

pin8-

EXT

RESET_L

pin9-

+5VDC

OUT

The FAIL and HIGH ALARM digital signal outputs are open drain 2N7002

outputs, able to sink about 50 mA each.

Detector Setups

Typical response and set points for the Product with NaI(TI) Scintillation

Detectors are as follows:
• Operating Voltage: 600 V dc to 1200 V dc determined by comparing plateaus
• Threshold: 10 mV dc (using a 39-inch cable)
• Calibration Constant: 2400 cpm/μR/hr
• Dead Time Correction: 5 µsec
• Linear Range with DTC: 10 µR/hr to 1500 µR/hr
Typical Checkpoints:
100

µR/hr

150 µR/hr

- calibration constant set point

200

µR/hr

500

µR/hr

1000 µR/hr

- dead time correction set point

1500

µR/hr