Operator Controls and Setup

Calibration Controls

Remove the calibration cover to expose the calibration controls. The calibration

controls include the up/down buttons, five calibration potentiometers, and the

option dipswitch (detailed in the following subsection). The five potentiometers

are detailed below:

Warning

To prevent possible electrical shock, fire, or personal injury:

To prevent contact with internal hazardous live parts that

are accessible using a tool: turn off the Product and
disconnect the power cord. Allow the Product to sit for
1 minute before accessing internal components.

Do not touch the circuit board in the calibration window

due to potential for electric shock.

ANALOG: Used to adjust the logarithmic analog voltage output. Adjusted in

calibration mode to the full-scale voltage reading or adjusted to a known point

at some given reading.

HV: Used to set the high voltage required for detector operation. Adjustable

from 0 V dc to 2500 V dc. The high voltage required will depend on the type

of detector used. Internal GM detectors typically require 550 V dc. Be sure to

check the high voltage with a high impedance (1000-Mohm impedance)

voltmeter only. A high-voltage checkpoint is located next to the HV

potentiometer.

DISC: Internal discriminator used to set negative pulse threshold for counting

pulses from the detector. Pad allows direct measurement of threshold

voltage. Utilize a Ludlum Model 500 Pulser or equivalent to inject pulses of

the desired threshold size. The pulse height threshold is adjustable from

2.0 mV dc to 100 mV dc.

BAT CHARGE: Used to set the backup battery trickle charging voltage. It is

set to 6.9 V dc while the battery is disconnected.

OVERLOAD: Used to set the detector current overload point. When

excessive radiation causes the detector to overload, this set point will cause

the FAIL light to engage, and the display will be forced to -OL-.