Common Options and Modifications

Time and Date Stamp Option

Description

When an alarm or failure occurs, the Product will print the current reading, date,

time, and either ALARM or FAIL to the RS-232 port. The instrument will print

once every 30 seconds as long as the alarm or fail condition is present.

Setup

You will need the following: the Product, a 1220 40-column printer, and a cable

(LMI PN: 8303-674)
The printer should be configured at 2400 bps (baud), no parity, 8 data bits, 1 stop

bit, and no handshaking. See printer manual for proper setup instructions.

Date and Time

Check the month and day (MMDD) by pressing the LOW ALARM and
HIGH ALARM

buttons simultaneously. The month and day will be displayed as

long as those buttons are pressed. The month and day can be set from 0101 to

1231.
Check the year (YYYY) by pressing the LOW ALARM and CAL CONST buttons

simultaneously. The year will be displayed as long as those buttons are pressed.

The year can be adjusted from 0000 to 9999.
Check the hours and minutes (HHMM) by pressing the LOW ALARM and
DEAD TIME

buttons simultaneously. The hours and minutes will be displayed as

long as those buttons are pressed. The hours and minutes can be adjusted from

0000 to 2359.

RS232 Data Format

The data will be sent to the RS-232 port as:

Byte

1

0 x Byte

18

Space

(20H)

Byte

2

x x Byte

19

H

Byte 3

x OR

x

Byte 20 H

Byte

4

x x Byte

21

:

Byte

5

. . Byte

22

M

Byte

6

x 0 Byte

23

M

Byte 7

Space (20H)

Byte 24 :

Byte 8

Space (20H)

Byte 25 S

Byte 9

Space (20H)

Byte 26 S

Byte

10

M

Byte

27

Space

(20H)

Byte

11

M

Byte

28

A

Space

Byte

12

/

Byte

29

L

F

Byte

13

D

Byte

30

A OR A

Byte

14

D

Byte

31

R

I

Byte

15

/

Byte

32

M

L

Byte

16

Y

Byte

33

Carriage Return (0DH)

Byte

17

Y

Byte

34

Line Feed (0AH)

Example output:

0642.1 04/21/95 16:56:24 ALARM
0000.0 04/21/95 08:32:16 FAIL

Modifications to the Product for Optimum Performance

The Product main board (LMI PN: 5396-160) has these modifications:
• U531 changes from an LM358 to an OPA2343UA; LMI PN: 06-6582
• C531 changes from 10 μF tantalum to 0.047μF “poly film” (Polypropylene

sulfide), LMI PN: 04-5729.

• R432 changes from 100 k to 1 mg, LMI PN: 12-7844