Introduction – Fluke Biomedical 05-450 User Manual

Page 9

Advertising
background image

1

Introduction

Model 05-450 and 05-450-2200 Digital Doorway Monitor (the Product) is

designed to monitor for nuclear radiation. Featuring a wall-mount chassis, the

Product has a four-digit LED display that is readable from 9 m (30 ft) away.

Backlit indicators warn of low radiation alarm (yellow), high radiation alarm (red),

instrument failure (red) and low battery (yellow). A green status light is a positive

indication of instrument operation.
Parameters are protected under a calibration cover. Calibration is easily

accomplished by moving the CAL dipswitch to the right, and using the

pushbuttons to increment or decrement the calibration constant, dead time

correction, and alarm point parameters. Parameters are stored in non-volatile

memory (retained even with power disconnected).
A five-decade logarithmic analog output is provided. A battery backup provides

48 hours of additional use after the primary power is removed.

Note

The detector does not contain any consumable materials.

Note

If the detector is used in a manner not intended by the manufacturer,
the detector may not function properly.

The Product has two scintillation detectors, each with internal lead shield to

reduce background radiation. Detectors may be supplied with or without

environmental enclosures. The scintillation detectors detect low levels of waste

radiation. These detectors are usually, but not always, used in scrap yards or

hospitals. Lead shields around the crystal allows specific coverage areas for

radiation detection. Figure shows the Product.