Radiation units, Checking parameters – Fluke Biomedical 05-450 User Manual

Page 16

Advertising
background image

05-450

Users Manual

8

Please set the alarm point(s) on this insturment
to conform to your requirements. The
factory-set alarm points may be incorrect for
your use.

Refer to the instrument manual for more
information on setting alarm points.

FAILURE TO RESET THE ALARM POINT(S)

MAY RESULT IN EXCESSIVE ALARMS OR

LACK OF SENSITIVITY.

huk02.eps

Figure 3. Alarm Point Notice

Initial power-up will momentarily activate the internal front-panel lights, sound the

audio, and display "8888" on the 4-digit LED display. The firmware version

number (39665Nyy) is then displayed as "396" and "65yy" (where yy represents

the current version number).
When the instrument has finished measuring background, it will display the

current radiation reading and begin checking for an alarm condition.

Radiation Units

The Product may be calibrated for almost any desired radiation units of measure.

Common units of measure include mR/hr, µR/hr, R/hr, mSv/h, µSv/h, cps, cpm,

and kcpm. In each case, the unit of measure is indicated underneath the four-

digit display. Throughout the rest of this manual, the notation <units> will be used

as a substitute.

Checking Parameters

Check the low alarm point setting by pressing the LOW ALARM button. The low

alarm point will be displayed as long as the button is pressed. The low alarm

point is in units of <units>. The low alarm point can be set from 0.1 <units> to

9999 <units>.
Check the high alarm point setting by pressing the HIGH ALARM button. The

high alarm point will be displayed as long as the button is pressed. The

highalarm point is in units of <units>. The high alarm point can be set from 0.1

<units> to 9999 <units>.
Check the calibration constant by pressing the CAL CONST button. The

calibration constant will be displayed as long as the button is pressed. The

calibration constant is in units of cpm (counts per minute) per <units>*. The

calibration constant can be set from 0.1 cpm/<units> to 9999 cpm/<units>.
Check the detector dead time correction by pressing down on the DEAD TIME

button. The dead time correction will be displayed as long as the button is

pressed. The dead time correction is in units of microseconds*. The dead time

correction can be set from 0.1 microseconds to 9999 microseconds.