Specifications, Line voltage, Load current – Fluke Biomedical ESA180 User Manual

Page 20

Advertising
background image

180

Users Guide

1-10

Table 1.2. Components and controls of the Analyzer (cont.)

K

Chassis
Connector

Provides a means for inputting the chassis
cable with its clip for connection to the DUT
chassis or enclosure. The chassis ground
resistance is measured with the FUNCTION
switch in the RESISTANCE position, and the
chassis leakage current is measured in the
CHASSIS position.

L

Dual Connector

Used to make point-to-point measurements
with optional leads available for purchase. For
leakage and voltage measurements, the black
cable with the clamp with black insulation is
attached to the CHASSIS connector, and the
black cable with the clamp with red insulation to
the DUAL connector. For point-to-point
resistance measurements, the black cables
with clamps with black insulation are attached
to the DUAL and CHASSIS connectors.

Specifications

The following are specifications for the Analyzer. Please contact your Fluke
Biomedical service representative for more information regarding the device
specifications.

Line Voltage

Range .....................................................................90 – 240 V ac 50/60 Hz
Accuracy.................................................................

±

3 % of reading,

±

1 LSD

Load Current

Range .....................................................................0 – 19.99 A
Accuracy.................................................................

±

5 % of reading,

±

1 LSD

Advertising
This manual is related to the following products: