Ground resistance, Leakage current, Isolation test – Fluke Biomedical ESA180 User Manual

Page 21: Current capacity, Current consumption, Environmental

Advertising
background image

Introduction and Specifications

Specifications

1

1-11

Ground Resistance

Range .................................................................... 0 – 2.00

Ω

Accuracy ................................................................

±

1 % of reading,

±

1 LSD

Range .................................................................... 2.01 – 19.99

Ω

Accuracy ................................................................

±

3 % of reading,

±

1 LSD

Current source ....................................................... 10 mA dc
Type....................................................................... 4-wire bridge

Leakage Current

Range .................................................................... 0 – 1999

μ

A

Accuracy:

DC and 25 Hz to 1 kHz ......................................

±

1 % of reading,

±

3 LSD

1.0 kHz to 100 kHz ............................................

±

2.5 % of reading,

±

3 LSD

100 kHz to 1 MHz ..............................................

±

5 % of reading,

±

3 LSD

Type measurement............................................ True rms
Impedance ......................................................... 1000

Ω

, AAMI ESI 1993

Isolation Test

Voltage................................................................... 110 % Line voltage

±

5%

Current................................................................... limited by 120 k

Ω

resistor

Current Capacity

Line 90 to 140 V ac................................................ 20 A max. on time 5 min.

15 A max. on time 30 min.

Line over 140 V ac................................................. 10 A

Current Consumption

ESA 180................................................................. 0.1 A @ 120 V – 60 Hz

Environmental

Operating Temperature ......................................... 15 – 40

°

C

Storage Temperature............................................. -20 – 65

°

C

Relative Humidity................................................... 90 % max.
Mains voltage range .............................................. 90 to 240 V ac