Measuring leakage current, Ground leakage current – Fluke Biomedical ESA180 User Manual

Page 30

Advertising
background image

180

Users Guide

2-8

Note

The OUTLET switch should be in the OFF (center) position for this
measurement.

Measuring Leakage Current

Leakage current is the flow of current through or over the surface of an
insulating material or insulator. For example, if a person comes in contact with
an operational device, leakage current is the amount of current that flows from
the point where the person came in contact with the device product, through
the person’s body, and back to ground. Measurement of leakage current is
required on all mains power products.

Ground Leakage Current

Ground Leakage Current [I

E

] (internal chassis current) is the current that flows

in the ground wire of the power cable to return the chassis leakage current to
true earth ground. It is only applicable to devices utilizing three-wire
(grounded) power cords.

This current does not constitute a hazard as long as the ground wire remains
intact and the current does not become excessive. Leakage current is due to the
proximity of the hot wire or line potential components to the chassis
represented by Z

L

, a combination of capacitance, C

L

, and resistance, R

L

,

components.

Note

The connection is made internally in the Analyzer, so no external
connectors are required for this test.

Use the following steps to measure ground leakage current:

1. Set the

FUNCTION

switch to

GROUND

. The leakage current is

displayed to 1999

μ

A (microamperes).