Fluke Biomedical ESA180 User Manual

Page 33

Advertising
background image

Operation, Maintenance, and Service

Measuring

Leakage

Current

2

2-11

Use the following steps to measure lead-to-ground current:

1. Connect the patient leads to the corresponding snaps on the top of the

Analyzer. If the device has 10 leads, connect the limb leads and C1 (V1)
initially and repeat with C2 (V2) through C6 (V6). Lead nomenclature for
this test is not important.

2. Select

LEAD-GND

with the

FUNCTION

switch and read leakage current

in microamperes for any combination of the patient lead selected by the

LEAD

switch.

V

V
A

A

µA

µA

L
E
A
K
A
G
E

- CURRENT

- CURRENT

- RESISTANCE

- RESISTANCE

- GROUND

- GROUND

- CHASSIS

- CHASSIS

- LEAD -GND

- LEAD -GND

- LEAD -LEAD

- LEAD -LEAD

- LEAD ISO

- LEAD ISO

- DUAL

- DUAL

- LINE VOLTS

- LINE VOLTS

- CURRENT

- CURRENT

- RESISTANCE

- RESISTANCE

- GROUND

- GROUND

- CHASSIS

- CHASSIS

- LEAD -GND

- LEAD -GND

- LEAD -LEAD

- LEAD -LEAD

- LEAD ISO

- LEAD ISO
- DUAL

- DUAL

fat08.eps

3. Make measurements under all combinations of the

OUTLET

switch,

NORMAL

and

REVERSE

; the

LIFT

GROUND

switch

CLOSED

and

OPEN

; and the device power turned

ON

and

OFF

.

4. Rotate the

LEAD

switch to each lead to test individually and then to

ALL

for testing with all leads connected together.

LEAD

LEAD

c

ra

ll

rl

rl

la

la

all

all

fat17.eps

Current measured should be the same for all leads, including the

ALL

position, as the current represents the isolation impedance to the patient
circuit.