Operation, Maintenance, and Service

Preliminary

Steps

2

2-3

Preliminary Steps

The Analyzer measures electrical parameters, as described below. These
voltages and currents are a natural phenomenon, and their presence within
reasonable limits does not constitute a hazard. However, it is necessary to
measure their values to determine if there is a significant change from previous
measurements or from the device specifications.

Figure 2-1 is a block diagram of a typical line-operated instrument with patient
connections. All measured parameters are labeled, and the following topics are
keyed to the labels in the diagram.

Hot

N

e

u

tral

Gro

u

nd

Earth

Enclos

u

re

Chassis

120/220

V

AC

Gro

u

nded

Data and

Control

Circ

u

its

Isolated

Patient

Circ

u

its

I

E

I

L

Z

I

R

G

R

L

V

L

C

L

C

L

Z

L

I

C

I

A

I

I

I

P

fat02.eps

Figure 2-1. Block Diagram of a Typical DUT

Preparing the Analyzer for Use

WX

Warning

To avoid possible electric shock, burning of the skin, or
personal injury to the patient, disconnect all patient
connections to the device before starting the following
procedure.