Measuring device current – Fluke Biomedical ESA180 User Manual

Page 28

Advertising
background image

180

Users Guide

2-6

device being tested. Depending on the device's operating current and the
electrical supply wiring, the voltage differential may be significant.

3. Check the value under load against the device's ratings to ensure that the

actual value remains within prescribed limits. An excessive drop also
suggests that a dedicated line of increased capacity should be run to the
instrument.

Measuring Device Current

Device, or Instrument, Current [I

L

] is the current used by the DUT. When

turned on, the device should be operated in its various modes to determine the
worst condition to track. Verify that the current of the DUT is within the
current rating of the Analyzer being used.

Measurement is made in the

HOT

wire via transformer coupling to ensure that

the total current is measured, as it is possible that the

NEUTRAL

and

GROUND

wire could share the return path.

Use the following steps to measure device current:

1. Switch the

FUNCTION

switch to

A - CURRENT

. The meter displays the

device's current to 19.99 A.

V

V
A

A

µA

µA

L
E
A
K
A
G
E

- CURRENT

- CURRENT

- RESISTANCE

- RESISTANCE

- GROUND

- GROUND

- CHASSIS

- CHASSIS

- LEAD -GND

- LEAD -GND

- LEAD -LEAD

- LEAD -LEAD

- LEAD ISO

- LEAD ISO

- DUAL

- DUAL

- LINE VOLTS

- LINE VOLTS

- CURRENT

- CURRENT

- RESISTANCE

- RESISTANCE

- GROUND

- GROUND

- CHASSIS

- CHASSIS

- LEAD -GND

- LEAD -GND

- LEAD -LEAD

- LEAD -LEAD

- LEAD ISO

- LEAD ISO
- DUAL

- DUAL

fat04.eps

2. Place the OUTLET switch in the NORMAL position; the NEUTRAL

switch remains in the CLOSED position.

3. Turn on the device and place it in its maximum load condition to obtain

the proper reading.

4. Log this data to note changes in values that indicate early problems.