ROUTINE CLEANING AND MAINTENANCE

The LifeChoice Activox POC was designed to minimize the amount of routine
maintenance that is required. The device does not have any end-user repairable
parts. Aside from the preventative maintenance outlined below, all other
maintenance required must be performed by qualified personnel.

Fan Inlet Vent Cleaning: There is one inlet vent located on the right side (facing
the unit) of the POC. The carrying case for the unit has mesh material corresponding
to the location of the Fan Inlet Vent. You should visually check the carrying case to
ensure that there is no buildup of lint, hair or other materials that could obstruct the
flow of air into the unit’s fan inlet vent. Remove any material buildup using a soft brush.

At least once a month inspect the grille of the
intake vent and if it appears dirty, use a dry cotton
swab (Q-Tip or similar) to clean it. Never use
alcohol, solvents, or petroleum distillates for
cleaning.

Unit and Carrying Case Cleaning: If it is
necessary to clean the unit and/or carrying case,
use only warm water and a mild liquid dish
d e t e r g e n t . D a m p e n

a c l o t h i n t h e s o l u t i o n a n d c a r e f u l l y w i p e t h e
outside surfaces of the unit and/or carrying case.

CAUTION: DO NOT use alcohol, solvents, polishes or any oily
substances since they may be flammable.

Nasal Cannula Inspection: At least once a week, visually inspect the nasal
cannula. Make sure there are no kinks or obstructions. Replace as needed with a
standard 7 ft. (2 m) single lumen nasal cannula from your local medical equipment
supplier.

Reserve Oxygen Supply: Your equipment provider should provide or suggest an
alternative source of supplemental oxygen therapy in case there is a mechanical
failure or power outage.

Hour Meter: The Hour Meter function is for use by trained service personnel
only.

Electronic Reset: If the indicators on the Control Panel do not appear as
described on page 15, follow the steps below to reset the device to default
factory settings.
1. Turn off the device.

2. Push and hold down the Power Button until the audible tone stops (about 10

seconds). Release the button and all LEDs will flash once.

3. Allow approximately 15 seconds for the device to cycle through a short

sequence. During this time, the unit will power on and off.

4. Power the device on.
If this reset doesn’t resolve the issue, contact your equipment provider.