The LifeChoice Activox Portable Oxygen Concentrator Model XYC100B-

SPT/XYC100B-PRO is intended for use in an electromagnetic environment

in which radiated RF disturbances are controlled. The customer or the user

of the Model XYC100B-SPT/XYC100B-PRO can help prevent electromagnetic

interference by maintaining a minimum distance between portable and mobile RF

communications equipment (transmitters) and the Model XYC100B-SPT/XYC100B-

PRO as recommended below, according to the maximum output power of the

communications equipment.

For transmitters rated at a maximum output power not listed above, the recommended

separation distance ‘d’ in meters can be estimated using the equation applicable

to the frequency of the transmitter, where ‘P’ is the maximum output power rating of the

transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80MHz and 800MHz, the separation distance for the higher frequency range

applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation

is affected by absorption and reflection from structures, objects, and people.

Recommended Separation Distances Between Portable and Mobile

RF Communications Equipment and The LifeChoice

®

Activox Portable

Oxygen Concentrator Model XYC100B-SPT/XYC100B-PRO

The LifeChoice Activox Portable Oxygen Concentrator Model XYC100B-SPT/

XYC100B-PRO is intended for use in the electromagnetic environment specified

below. The customer or the user of the Model XYC100B-SPT/XYC100B-PRO

should assure that it is used in such an environment.

Guidance and Manufacturer’s Declaration - Electromagnetic Emissions

Separation Distance According to the Frequency of Transmitter (m)

0.01

0.12

0.12

0.23

0.1

0.37

0.37

0.74

1

1.2

1.2

2.3

10

3.7

3.7

7.4

100

12

12

23

Rated Maximum

Output Power of

Transmitter (W)

150kHz to 80MHz

80MHz to 800MHz 800MHz to 2.5GHz

d=1.17 √P

d=1.17 √P

d=2.33 √P

Emissions Test

Compliance

Electromagnetic Environment - Guidance

The Model XYC100B-SPT/XYC100B-PRO uses RF

energy only for its internal function. Therefore, its RF

emissions are very low and are not likely to cause

any interference to nearby electronic equipment.

RF Emissions, CISPR 11

RF Emissions, CISPR 11

Harmonic Emissions,

IEC 61000-3-2

Voltage Fluctuations/Flicker

Emissions, IEC 61000-3-3

Group 1

Class B

Class A

Complies

The Model XYC100B-SPT/XYC100B-PRO is

suitable for use in all establishments, including

domestic establishments and those directly

connected to the public low-voltage power supply

network that supplies buildings used for domestic

purposes.