The LifeChoice Activox Portable Oxygen Concentrator Model XYC100B-SPT/
XYC100B-PRO is intended for use in the electromagnetic environment specified
below. The customer or the user of the Model XYC100B-SPT/XYC100B-PRO
should assure that it is used in such an environment.

NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic

propagation is affected by absorption and reflection from structures, objects, and

people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV

broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic

environment due to fixed RF transmitters, an electromagnetic site survey should be

considered. If the measured field strength in the location in which the Model XYC100B-

SPT/XYC100B-PRO is to be used exceeds the applicable RF compliance level above,

the Model XYC100B-SPT/XYC100B-PRO should be observed to verify normal

operation. If abnormal performance is observed, additional measures may be

necessary, such as re-orienting or relocating the Model XYC100B-SPT/XYC100B-

PRO.
b Over the frequency range 150kHz to 80MHz, field strengths should be less than 3 V/m.

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

Immunity Test

IEC 60601

Compliance

Electromagnetic

Test Level

Level

Environment - Guidance

Portable and Mobile RF

communications equipment

should be used no closer

to any part of the Model

XYC100B-SPT/XYC100B-PRO,

including cables, than the

recommended separation

distance calculated from the

equation applicable to the

frequency of the transmitter.

Recommended

Separation Distance

d = 1.17 √P

d = 1.17 √P, 80MHz

to 800MHz

d = 2.33 √P, 800MHz to

2.5GHz

3 Vrms

150kHz-80MHz

3 V/m

80MHz-2.5GHz

3 V

3 V/m

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

Where ‘P’ is the maximum output power rating of the transmitter in watts (W)

according to the transmitter manufacturer and ‘d’ is the recommended separation

distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic

site survey,

a

should be less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with the following

symbol: