Appendix – Ivoclar Vivadent Bluephase Meter User Manual

Page 35

Advertising
background image

Appendix

Emission test

RF emissions CISPR 11

RF emissions CISPR 11

Harmonic emissions IEC 61000-3-2

Voltage fluctuations / flicker emissions
IEC 61000-3-3

Compliance

Group 1

Class B

No power network input

No power network input

Electromagnetic environment - guidance

The “Bluephase Meter” uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic equipment.

The “Bluephase Meter” is suitable for use in all establishments, including domestic establish-
ments and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.

Immunity test

Electrostatic
discharge (ESD)
IEC 61000-4-2

Electrical fast
transient/burst
IEC 61000-4-4

Surge
IEC 61000-4-5

Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11

Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8

IEC 60601 test level

± 6 kV contact
± 8 kV air

± 2 kV for power supply lines
± 1 kV for input/output lines

± 1 kV line - line
± 2 kV line - earth

<5 % U

T

(>95 % dip in U

T

) for 0.5 cycle

40 % U

T

(60 % dip in U

T

) for 5 cycles

70 % U

T

(30 % dip in U

T

) for 25 cycles

<5 % U

T

(>95 % dip in U

T

) for 5 sec

3 A/m

Compliance level

± 6 kV contact
± 8 kV air

n.a. (no cables connected)

n.a. (no cables connected)

n.a. (no mains power)

30 A/m

Electromagnetic environment - guidance

Floors should be concrete or ceramic tile. If floors are covered with synthetic material,
the relative humidity should be at least 30%.

Power frequency magnetic fields should be at levels characteristic of a typical location
in a typical commercial or dental environment.

Guidance and manufacturer's declaration - electromagnetic emission

The “Bluephase Meter” is intended for use in the electromagnetic environment specified below. The customer or the user of the “Bluephase Meter”
should assure that it is used in such an environment.

Guidance and manufacturer's declaration - electromagnetic immunity

The “Bluephase Meter” is intended for use in the electromagnetic environment specified below. The customer or the user of the “Bluephase Meter”
should assure that it is used in such an environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.

Bluephase Meter-BA-WE3_611504_Rev4_bluephase meter.qxd 24.10.12 08:12 Seite 35