K-Patents PR-33-S User Manual

Page 32

Advertising
background image

28

PR-33-S instruction manual

Document/Revision No. Rev. 1.4

Effective: May 15, 2015

4. The light source is faulty. When the sensor is removed from the process, the yellow

flashing light can be seen through the prism.

Note: The light is only visible at an oblique angle. Also check the LED value on the

Diagnostics page (section 2.4.3); if the value is clearly below 100, LED fault is not

likely.

5. There are negative spikes in the optical image. The probable cause is dust or finger-

prints on the CCD window. Please contact K-Patents.

6. The CCD card in the sensor is faulty. Please contact K-Patents.

7.2.3 Message

PRISM COATED

Cause: The optical surface of the prism is coated by the process medium or impurities

in the process medium.

Action: Remove sensor from line and clean prism manually.

If the problem is recurrent, consider improving the flow conditions (see Section 3.1.1,

“Choosing sensor mounting location”).

7.2.4 Message

LOW IMAGE QUALITY

Cause: The most likely cause for this message is coating on the prism. There still is a

optical image available, but the measurement quality may not be optimal.

Action: Clean the prism, see Section 7.2.3 above.

7.2.5 Message

NO SAMPLE

The operation of the equipment is OK but there is no process liquid on the prism. In

some cases this message may also be caused by coating on the prism.

7.2.6 Message

TEMP MEASUREMENT FAULT

Indicates faulty temperature element. Please contact K-Patents.

Note: A difference to some other process temperature measurement is not a fault.

The Pr-33-S shows the true temperature of the prism surface.

Advertising