7 Troubleshooting

29

© Copyright K-Patents 2015. All rights reserved.

7.2.7 Concentration drift during

NORMAL OPERATION

For drift upward or downwards to zero, suspect prism coating and clean the prism.

7.3 Diagnostic messages table

Important:

The messages are listed in descending order of priority. Example: If both

NO OPTICAL IMAGE

and

TEMP MEASUREMENT FAULT

are activated, only

NO OPTICAL IMAGE

will be

shown.

Message

Section

OUTSIDE LIGHT ERROR

7.2.1

NO OPTICAL IMAGE

7.2.2

TEMP MEASUREMENT FAULT

7.2.6

HIGH SENSOR HUMIDITY

7.1.1

HIGH SENSOR TEMP

7.1.2

NO SAMPLE

7.2.5

PRISM COATED

7.2.3

OUTSIDE LIGHT TO PRISM

7.2.1

LOW IMAGE QUALITY

7.2.4

NORMAL OPERATION