Microscope Software manual http://www.ViTiny.com

60

Copyright © 2008-2014 MicroLinks Technology Corp.

ratio display

 To delete and rename scale file
 Choose delete or rename the file from the drop down menu, see Fig.5-59.

Then click “delete” or “rename”.

Fig.5-59 drop down menu choose file

 Semi-Auto Calibration

Set up a customized semi-auto Calibration. See Fig. 5-60. “Input

Measurement Length” means the real length. For example, enter 400 means

4.00mm, enter 50 means 0.5mm. The Display Scale check-box is to decide to
show the Display Scale or not. If click the Horizontal Fixed, only a horizontal axis

can draw on the image; on the contrary, if click the Vertical fixed, only a vertical

axis can draw on the image. If choose “None”, axis can draw from any direction. A

horizontal or vertical axis can be draw by pressing ‘shift’ button on your keyboard
together with clicking mouse left. After it is completed, please click “Finished” for

saving the new scale.

Fig. 5-60 Input the actual measurement length

The calibration frame can be freeze by pressing the right button on the

mouse. The freeze frame helps for calibrating in a stable condition. Put the

calibrator under the machine and adjust focus to clearly see the scale. Key in

“400” length in the check-box of “Input Measurement Length”. Use mouse to

draw a yellow line for 4.00mm.Click “Finished” to complete and save the new

calibration figure.

The new setting scale (Calibration Scale [New]) show on the frame. See