Configure:fresistance, Configure:frequency – RIGOL DM3058 User Manual
Page 107

RIGOL
Programming Guide for DM3058/DM3058E
4-14
Explanations:
The unit is Ω.
For <range>: MIN = 200; MAX = 100 000 000, DEF=AUTO.
<resolution> relates to both the measurement rate and integration time, see
Table 4-2.
6. CONFigure:FRESistance
Command Format:
CONFigure:FRESistance
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Function:
The command presets the multimeter as specified range and resolution for 4-wire
resistance measurement, but does not start the measurement.
Parameter:
Name
Type
Range
Default
<range>
Discrete
{200 Ω|2 kΩ|20 kΩ|200 kΩ
|2 MΩ|10 MΩ|100 MΩ|AUTO}
AUTO
<resolution> Discrete
Refer to Table 4-2
1ppm x range
Explanations:
The unit is Ω.
For <range>: MIN = 200; MAX = 100 000 000, DEF=AUTO.
<resolution> relates to both the measurement rate and integration time, see
Table 4-2.
7. CONFigure:FREQuency
Command Format:
CONFigure:FREQuency
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Function:
The command presets the multimeter as specified range and resolution for