Sense:]voltage:ac:resolution – RIGOL DM3058 User Manual
Page 135

RIGOL
Programming Guide for DM3058/DM3058E
4-42
Function:
The commands set and query the precision of DC voltage measurement.
Parameter:
Name
Type
Range
Default
<resolution> Discrete
Refer to Table 4-2
1ppm x range
Explanations:
[SENSe:]VOLTage:DC:RESolution {<resolution>|MINimum|MAXimum} – Sets
the desired precision of DC voltage measurement.
[SENSe:]VOLTage:DC:RESolution? [MINimum|MAXimum] – Queries the current
precision of DC voltage measurement.
<resolution> relates to measurement rate and integration time, refer to Table
4-2.
Return Format:
The query returns the current precision of DC voltage measurement.
19. [SENSe:]VOLTage:AC:RESolution
Command Format:
[SENSe:]VOLTage:AC:RESolution {<resolution>|MINimum|MAXimum}
[SENSe:]VOLTage:AC:RESolution? [MINimum|MAXimum]
Function:
The commands set and query the precision of AC voltage measurement.
Parameter:
Name
Type
Range
Default
<resolution> Discrete
Refer to Table 4-2
1ppm x range
Explanations:
[SENSe:]VOLTage:AC:RESolution {<resolution>|MINimum|MAXimum} – Sets the
desired precision of AC voltage measurement.
[SENSe:]VOLTage:AC:RESolution? [MINimum|MAXimum] – Queries the current
precision of AC voltage measurement.
<resolution> relates to measurement rate and integration time, refer to Table