Measure:voltage:dc – RIGOL DM3058 User Manual
Page 53

RIGOL
Programming Guide for DM3058/DM3058E
3-28
The command returns the measured DC voltage value in scientific notation, such
as: 8.492853e-05, the unit is V.
4. :MEASure:VOLTage:DC
Command Format:
:MEASure:VOLTage:DC {<range>|MIN|MAX|DEF}
Function:
The command sets the range and resolution of DC voltage measurement.
Parameter:
Name
Type
Range
Default
<range>
Discrete
{0|1|2|3|4}
2
Explanations:
The corresponding resultion will be set automatically after you set the range.
The measurement type will change into “Manual” automaticlly as you set the
range.
For the parameter range and the corresponding resolution, see table below.
Table 3-4 DC voltage range and
the
corresponding resolution
Parameter
Range
Resolution
0
200 mV
100 nV
1
2 V
1 μV
2
20 V
10 μV
3
200 V
100 μV
4
1000 V
1 mV
MIN
200 mV
100 nV
MAX
1000 V
1 mV
DEF
20 V
10 μV
Example:
Sets the DC votage range to 200V.
:MEASure:VOLTage:DC 3
Sets the DC votage range to the minimum.