Sense:]resistance:range – RIGOL DM3058 User Manual

Page 127

Advertising
background image

RIGOL

Programming Guide for DM3058/DM3058E

4-34

Function:
The commands set and query the range of AC current measurement.

Parameter:

Name

Type

Range

Default

<range> Discrete

{ 20 mA|200 mA|2 A|10 A }

200 mA


Explanations:
[SENSe:]CURRent:AC:RANGe {<range>|MINimum|MAXimum} - Sets the desired
range of AC current measurement from {20 mA|200 mA|2 A|10 A }; thereinto,
MIN = 20 mA, MAX = 10 A and the unit is A. If a 200 mA range is required, enter
0.2 as its parameter.
[SENSe:]CURRent:AC:RANGe? [MINimum|MAXimum] – Queries the current
range of AC current measurement.

Return Format:
The query returns the range of AC current measurement.


6. [SENSe:]RESistance:RANGe

Command Format:
[SENSe:]RESistance:RANGe {<range>|MINimum|MAXimum}
[SENSe:]RESistance:RANGe? [MINimum|MAXimum]

Function:
The commands set and query the range of 2-wire resistance measurement.

Parameter:

Name

Type

Range

Default

<range>

Discrete

{200 Ω|2 kΩ|20 kΩ|200 kΩ

|2 MΩ|10 MΩ|100 MΩ}

200 kΩ


Explanations:
[SENSe:]RESistance:RANGe {<range>|MINimum|MAXimum} - Sets the desired
range of 2-wire resistance measurement from {200 Ω|2 kΩ|20 kΩ|200 kΩ|2
MΩ|10 MΩ|100 MΩ}; thereinto, MIN = 200 Ω, MAX = 100 MΩ and the unit is Ω. If
a 2 kΩ range is required, enter 2000 as its parameter.