Format commands, Format commands -23 – RIGOL DM3058 User Manual
Page 172
Advertising

RIGOL
Programming Guide for DM3058/DM3058E
5-23
Format Commands
1. FORMAT
Command Format:
FORMAT <frmt>
FORMAT?
Function:
The commands set and query the output format.
Explanations:
<frmt> could be 1 or 2. 1 denotes a measured value without unit, 2 permits the
unit following the value. For the relationship of each function and unit, see table
below:
Table 5-3 Units of format 2
Function
Unit
DC voltage
VDC
AC voltage
VAC
DC current
ADC
AC current
AAC
Resistance
OHMS
Frequency
Hz
Diode
VDC
Continuity
OHMS
Note: Being restricted by working principle, command FORMAT <frmt> and
FORMAT? are only responsed without practical operation.
Advertising