Sense:]function, Sense:]voltage:dc:range – RIGOL DM3058 User Manual
Page 124

RIGOL
Programming Guide for DM3058/DM3058E
4-31
1. [SENSe:]FUNCtion
Command Format:
[SENSe:]FUNCtion "<function>"
[SENSe:]FUNCtion?
Function:
The commands set and query the current measurement function.
Parameter:
Name
Type
Range
Default
<function> Discrete
Refer to Table 4-3
—
Explanations:
[SENSe:]FUNCtion "<function>" – Selects the measurement function to be used.
Note this comand has no default value and <function> must be set, see table
below.
[SENSe:]FUNCtion? – Queries the current measurement function.
Table 4-3 Parameters and explanations
Parameter
Meaning
Parameter
Meaning
VOLTage:DC
DC voltage
PERiod
Period
VOLTage:AC
AC voltage
RESistance
2WR
CURRent:AC
AC current
FRESistance
4WR
CURRent:DC
DC current
CONTinuity
Continuity
FREQuency
Frequency
DIODe
Diode
Return Format:
The query returns the current measurement function.
2. [SENSe:]VOLTage:DC:RANGe
Command Format:
[SENSe:]VOLTage:DC:RANGe {<range>|MINimum|MAXimum}
[SENSe:]VOLTage:DC:RANGe? [MINimum|MAXimum]
Function: