B. gas requirements, C. physical, Model nga2000 hfid – Emerson Process Management ROSEMOUNT NGA2000 HFID User Manual

Page 19: Instruction manual

Advertising
background image

Instruction Manual

748297-E

June 2003

Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications

1-5

Model NGA2000 HFID

b. Gas

Requirements

Sample .......................................... Non-flammable, below 100% of LEL

Flow rate ..............................1.0 to 2.5 L/min.
Supply pressure...................345 to 620 hPa-gauge (5 to 9 psig)
Temperature ........................110

o

C to 230

o

C (230

o

F to 446

o

F), <20

o

C variance/24 hours, <10

o

C

variance/hr.

Particulates ..........................filtered to <2 microns
Dewpoint..............................<15

o

C below the setpoint

Regulated Air................................. Instrument air or nitrogen

Flow rate ..............................2 to 4 L/min.
THC .....................................

≤2 ppm, CH

4

Supply pressure...................689 to 1723 hPa-gauge (10 to 25 psig)
Particulates ..........................filtered to <2 microns

Purge Air:....................................... Instrument air, nitrogen or other nonflammable gas (refer to

ANSI/NFPA 496 for the requirements for the Protective Gas Sys-
tem)

Flow rate: ............................16 to 18 L/min.
Supply pressure:..................689 to 1378 hPa-gauge (10 to 20 psig)

Fuel Gas ........................................ Premixed 40% hydrogen and 60% helium

Flow rate ..............................80 to 100 ml/min
THC .....................................

≤0.5 ppm, CH

4

Supply pressure...................3101 to 3446 hPa-gauge (45 to 50 psig)

WARNING

EXPLOSION HAZARD

Do not use pure hydrogen fuel. An explosion resulting in severe personal injury or death could oc-
cur.

Burner Air ...................................... Zero-grade air

Flow rate ..............................350 to 400 mL/min.
THC .....................................

≤1 ppm, CH

4

Supply pressure...................1723 to 3446 hPa-gauge (25 to 50 psig)

c. Physical

Case Classification ........................ General purpose for installation in weather-protected area

Maximum Separation .................... 1600m (1 mile) from Analyzer Module to Platform

Materials in Contact With Sample . Stainless steel and glass-filled teflon

Dimensions.................................... See Outline and Mounting Dimensions, Figure 2-5 on page 2-7

Weight ........................................... 15.9 kg (35 lbs.)

Mounting........................................ Horizontally, custom-installed in a panel