4-16 Maintenance and Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA2000 HFID

Case Pressure Purge Switch

1. Per Section 4-2e on page 4-13, re-

move securing hardware from elec-
tronics assembly and lay on bench.

2. Disconnect the two electrical termi-

nals, note location.

3. Disconnect tube at pressure switch.

4. Remove mounting screws (2) and

washers (2).

Reverse procedure for installation of re-
placement switch.


































Figure 4-14.

Case Pressure Purge Switch Installation









The bracket does not have to be removed from the electronics assembly for this procedure.

Case Pressure Purge
Switch 903690