Instruction Manual

748297-E
June 2003

4-24 Maintenance and Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA2000 HFID

LON/Power Module

1. Disconnect wiring connectors, note

locations.

2. Refer to Figure 4-20 on page 4-22.

From the outside of the front panel,
remove the two mounting screws.

3. Install replacement module in reverse

order.

LED Indicator Assembly

1. Disconnect wiring connector, note lo-

cation.

2. Refer to Figure 4-20 on page 4-22.

From the inside of the front panel,
remove the two hex nuts securing
LED indicator assembly to front panel.
Remove indicator assembly and o-
rings (four).

3. Inspect o-rings for damage, replace if

necessary. Install o-rings on re-
placement indicator assembly, mount
assembly on mounting studs with hex
nuts.

4. Re-connect wiring connector.

Manual Ignite Toggle Switch

1. Disconnect wiring connector, note lo-

cation.

2. Refer to Figure 4-20 on page 4-22.

From the outside of the front panel,
remove the toggle switch seal.

3. Pull the switch and o-ring out from in-

side the front panel.

4. Inspect o-ring for damage, replace if

necessary. Install o-ring on replace-
ment switch, insert through front
panel from the inside.

5. Install

switch

seal.

6. Re-connect wiring connector.

Burner Air Sensor

1. Disconnect wiring connector, note lo-

cation.

2. Using an open-end wrench to hold the

sensor fitting while using another
open-end wrench to remove the sen-
sor.

3. Replace the Teflon pipe thread tape

on the treads of the sensor fitting.

4. Install sensor onto sensor fitting.

5. Re-connect wiring connector.

Fuel Sensor

1. Disconnect wiring connector, note lo-

cation.

2. Using an open-end wrench to hold the

sensor fitting while using another
open-end wrench to remove the sen-
sor.

3. Replace the Teflon pipe thread tape

on the treads of the sensor fitting.

4. Install sensor onto sensor fitting.

5. Re-connect wiring connector.