Model nga2000 hfid, Instruction manual – Emerson Process Management ROSEMOUNT NGA2000 HFID User Manual

Page 37

Advertising
background image

Instruction Manual

748297-E

June 2003

Rosemount Analytical Inc. A Division of Emerson Process Management Startup and Operation 3-7

Model NGA2000 HFID

0

0.2

0.4

0.6

0.8

1.0

RESPONSE

(100 ppm CH

4

fullscale)

SAMPLE: 100 ppm CH

4

in N

2

at 3.5 psig (241 hPa)

0

5

344

10

688

15

1032

20

1376

25

1726

30

2064

psig

hPa

FUEL PRESSURE

AIR: 30 psig (2064 hPa)

AIR: 20 psig (1376 hPa)

AIR: 10 psig (688 hPa)

0

0.2

0.4

0.6

0.8

1.0

RESPONSE

(100 ppm CH

4

fullscale)

SAMPLE: 100 ppm CH

4

in N

2

at 3.5 psig (241 hPa)

0

5

344

10

688

15

1032

20

1376

25

1726

30

2064

psig

hPa

AIR PRESSURE

FUEL: 30 psig (2064 hPa) H

2

FUEL: 25 psig (1726 hPa) H

2

FUEL: 20 psig (1376 hPa) H

2
























Figure 3-1.

Typical Curves of Module Response vs. Pressure Setting on Fuel Pressure Regulator


























Figure 3-2.

Typical Curves of Module Response vs. Pressure Setting on Air Pressure Regulator

Advertising