Maintenance and Troubleshooting 4-15

Model NGA2000 HFID

Case Temperature Sensor

1. Per Section 4-2e on page 4-13, re-

move securing hardware from elec-
tronics assembly and lay on bench.

2. Disconnect case temperature sensor

cable.

3. Remove screw securing cable clamp

holder to signal board.

4. Remove case temperature sensor

from cable clamp holder.

5. Per Figure 4-13 below insert re-

placement case temperature sensor
into cable clamp holder.

Re-assemble to signal board mounting
screw.




















Figure 4-13.

Case Temperature Sensor Installation




















Center

sensor in

cable clamp