Ignitor, Flameout sensor, Model nga2000 hfid – Emerson Process Management ROSEMOUNT NGA2000 HFID User Manual

Page 44: Instruction manual

Advertising
background image

Instruction Manual

748297-E
June 2003

4-6 Maintenance and Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA2000 HFID

The components shown can be replaced without removing burner/thermal block from oven. Oven not shown for clarity.
Thermal block shown in phantom for clarity.

Set Screw
M3X0.5 x 10mm
903125

RTD Detector
657063

Ignitor Assembly
657205

O-Ring
903736

O-Ring
903737

Flameout Sensor
657199

Temperature Sensor
657468

Retainer, Burner Cap

Burner Cap

Ignitor

1. Refer to Figure 4-4A on page 4-5.

Remove the four screws on the oven
cover, remove cover.

2. Refer to Figure 4-5 below. Discon-

nect the ignitor wiring connector, note
location.

3. Using an open-end wrench, unscrew

the ignitor assembly from the burner.
Verify that o-ring is also removed.

4. Install replacement ignitor and new o-

ring. Using open-end wrench, snug
down. Do not over-tighten!

5. Re-attach wiring connector.

6. Install oven cover

Flameout Sensor

1. Refer to Figure 4-4A on page 4-5.

Remove the four screws on the oven
cover, remove cover.

2. Refer to Figure 4-5 below. Discon-

nect the flameout detector wiring con-
nector, note location.

3. Lift up the burner cap until flameout

sensor is accessible. Using an open-
end wrench, unscrew the flameout
detector from the burner. Verify that
o-ring is also removed.

4. Install replacement flameout detector

and new o-ring. Using open-end
wrench, snug down. Do not over-
tighten!

5. Re-attach wiring connector.

6. Install oven cover.




























Figure 4-5.

Burner - Sensor, Flameout Detector, RTD Detector and Ignitor

Advertising