Maintenance and Troubleshooting 4-9

Model NGA2000 HFID

Air Baffle (Ref)
102260
See Detail

Sample Jet
657005

Gasket
102256

Jet Assembly
657012

Gasket
102273

Nut, Jet
657016


Air Baffle

≈ .06

Position with tab over
the air hole
(approximately 1/16” to
the right).

End bent to raise it above the tab a
distance equal to the height of the tab.

5. Holding burner base, lift burner cap

off of assembly, set aside.

6. Remove

gasket.

7. Refer to Figure 4-7B on page 4-8.

Holding burner base, lift combustion
chamber off, set aside.

8. Refer to Figure 4-8 below. Lift air baf-

fle out of burner base.

9. Remove the sample jet and gasket

from the bottom of the burner base.

10. Remove the jet nut. Grasp jet as-

sembly and lift out (along with upper
gasket) of burner base. Remove bot-
tom gasket.









































Figure 4-8.

Burner Jets