Amprobe ACD-30P Clamp-On-Power-Meters User Manual

Page 15

Advertising
background image

14

In

T

P

V
O

Update Rate :
Power function: 1 per second nominal
Voltage, ACA clamp-on, Ohm, Hz & Temperature functions: 4 per second nominal
Polarity : Automatic
Low Battery : Below approx. 2.4V
Operating Temperature : 0

o

C to 40

o

C

Relative Humidity : Maximum relative humidity 80% for temperature up to 31

o

C

decreasing linearly to 50% relative humidity at 40

o

C

Altitude : Operating below 2000m
Storage Temperature : -20

o

C to 60

o

C, < 80% R.H. (with battery removed)

Temperature Coefficient : nominal 0.15 x (specified accuracy)/

o

C @(0

o

C -18

o

C or

28

o

C -40

o

C), or otherwise specified

Sensing : True RMS sensing for all models
Safety : Meets IEC61010-2-032 (1994), EN61010-2-032(1995), UL3111-2-032(1999).

Measurement Category : III 600 Volts ac & dc

Transient protection : 6.5kV (1.2/50

µs surge) for all models

Pollution degree : 2

E.M.C. : Meets EN61326(1997, 1998/A1), EN61000-4-2(1995), and EN61000-4-3(1996)

an RF field of 3V/m:

otal Accuracy = Specified Accuracy + 45 digits

erformance above 3V/m is not specified

Overload Protections :

ACA Clamp-on jaws : AC 1000A rms continuous

+ & COM terminals (all functions) : 600VDC/VAC rms

Power Supply : standard 1.5V AAA Size (NEDA 24A or IEC LR03) battery X 2
Power Consumption :

oltage, ACA, Hz & Power functions: 10mA typical

hm & Temperature functions: 4mA typical

APO Timing : Idle for 17 minutes
APO Consumption : 10

µA typical

Dimension : L224mm X W78mm X H40mm
Weight : 224 gm approx
Jaw opening & Conductor diameter : 45mm max
Special features : Backlight display (model ACD-31P & ACD-41PQ only); AutoVA

TM

(Auto Selection on ACV, DCV or ACA functions); Power measurement of selectable W,
VAR & VA with dual-display Total Power Factor features; Total harmonic distortion
THD%-F (model ACD-41PQ only); PEAK-rms HOLD