Input warning

The meter beeps as well as displays “InErr” to warn the user against possible damage to
the meter due to improper connections to the

µ

A, mA, or A input jacks when other

function (like voltage function) is selected.

Line filter frequency selection

Press the 40000 button while turning the meter on to display the set frequency. Press the

SELECT button for 50Hz or press the STORE button for 60Hz to cope with your power

line frequency. Then press the RANGE button to store the selected frequency. Selecting

the appropriate line filter frequency can maximize the meter’s noise rejection ability.

Set Beeper Off

Press the SELECT button while turning the meter on to disable the Beeper feature. This

does not disable the input warning beeper.

5) MAINTENANCE

WARNING

To avoid electrical shock, disconnect the meter from any circuit, remove the test leads

from the input jacks and turn OFF the meter before opening the case. Do not operate

with open case. Install only the same type of fuse or equivalent

Cleaning and Storage

Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or

solvents. If the meter is not to be used for periods of longer than 60 days, remove the

battery and store it separately

Trouble Shooting

If the instrument fails to operate, check battery, fuses, leads, etc., and replace as

necessary. Double check operating procedure as described in this user’s manual

If the instrument voltage-resistance input terminal has subjected to high voltage transient

(caused by lightning or switching surge to the system) by accident or abnormal conditions

of operation, the series fusible resistors will be blown off (become high impedance) like

fuses to protect the user and the instrument. Most measuring functions through this