Amprobe AM91RS Digital-Multimeter User Manual
Page 28
27
Duty Cycle
RANGE
Accuracy
0.1%
99.9%
0.5d/kHz
2d
Input Frequency : 50Hz
250kHz; 5V Logic Family with + & - pulse width
2
µ
s
Update Rate : 1.2 per second nominal
Overload protection : 600VDC/VAC rms
Capacitance
RANGE
Accuracy*
4.000nF**
4.0% + 10d
40.00nF
3.0% + 5d
400.0nF
0.8% + 5d
4.000
µ
F
0.8% + 3d
40.00
µ
F
2.0% + 3d
400.0
µ
F
5.0% + 5d
4.000mF
5.0% + 5d
40.00mF
6.0% + 5d
*Accuracies with film capacitor or better
**Specified from 10% to 100% of range
Overload protection : 600VDC/VAC rms
Ohms
RANGE
Accuracy
40.00
Ω
0.2% + 6d
400.0
Ω
, 4.000k
Ω
, 40.00k
Ω
, 400.0k
Ω
0.15% + 2d
4.000M
Ω
0.3% + 2d
40.00M
Ω
1.5% + 5d
400.0nS
0.7% + 5d
Open Circuit Voltage :
1.3VDC (
3VDC for 40
Ω
& 400
Ω
ranges)
Temperature coefficient : 0.1 X (Specified accuracy)/
@ 23F - 65F or 83F - 122F (0
18
or 28
50
)
Overload protection : 600VDC/VAC rms
Diode Tester
Range
Accuracy
Test Current (Typical) Open Circuit Voltage
4.000V
2%+1d
0.8mA
3.5 VDC
Overload Protection : 600VDC/VAC rms
Audible Continuity Tester
Audible threshold : between 10
Ω
and 60
Ω
. Response time < 150
µ
s
Overload protection : 600VDC/VAC rms