Specifications – Amprobe HD160C Digital-Multimeter User Manual

Page 18

Advertising
background image

12

Specifications

General Specifications
Display: 4 digit LCD, 9999 counts, with

annunciators, menu features and 41 segment

bargraph.
Polarity Indication: Automatic
Input overload indication: OL, -OL.
Low Battery Indication: B; less than 50 hours

battery life remain, accuracy is no longer

guaranteed
Display Update Rate: 2/sec, nominal; 20/sec for

bargraph.
Oper. Temp. 0 °C to +50 °C @ 0 to 75 % R.H.
Storage Temp: -20 °C to 60 °C @ 0 to 80 % RH,

battery removed
Altitude: 2000 meters - indoor/outdoor use
Temperature coefficient: 0.1 x (spec. accuracy)/

°C (0 °C to 18 °C and 28 °C to 50 °C)
Fuse: F 2 A/1500 V fuse (8 mm x 65 mm),

I.R. 30 kA – Amprobe p/n FP700
Power: Standard 9-volt battery, NEDA 1604, JIS

006P, IEC 6F22
Auto Power-Down: Meter powers down after

approximately 30 minutes of inactivity. Not in

Min/Max function.
Battery Life (typical): 150 hours, alkaline.

Backlight usage consumes extra power and will

decrease battery life significantly. Backlight

auto-off after approx. 60 seconds.
Dimensions, without holster (H x W x D): 200

x102 x59 mm (7.9 in x 4. 0 in x 2.3 in)
Weight (incl. battery): 642 g (22 oz)
Accessories: Heavy Duty Test leads with

threaded alligator clips, battery (installed),

hex wrench (inside holster), Holster with

Magne-Grip strap, Type K thermocouple probe

(TP255A), Temperature Adaptor (TA-1A) and

User Manual.
Case material: Reinforced, high-impact, fire

retardant thermoplastic
Safety: Meets EN 61010-1 Cat IV -1500 V dc or

1000 V ac. Class II. EN 60529:IP67
EMC: Meets EN 61326-1

EMC: This product complies with

requirements of the following European

Community Directives: 89/336/EEC

(Electromagnetic Compatibility) and 73/23/EEC

(Low Voltage) as amended by 93/68/EEC (CE

Marking). However, electrical noise or intense

electromagnetic fields in the vicinity of the

equipment may disturb the measurement

circuit. Measuring instruments will also respond

to unwanted signals that may be present within

the measurement circuit. Users should exercise

care and take appropriate precautions to avoid

misleading results when making measurements

in the presence of electronic interference.

Electrical Specifications
Accuracy at 23 °C ± 5 °C, <75 % RH, guaranteed
for one year.

DC Volts
Ranges: 1000 mV, 10 V, 100 V, 1500 V
Resolution: 0.1 mV in 1000 mV range:
Accuracy: ±(0.1 % rdg + 5 dgt)
Input Impedance: 10 MΩ
CMRR: >120 dB up to 1500 V dc
NMRR: >60 dB at 50 or 60 Hz
OL Protection: 1500 V dc or 1000 V ac rms.
Transient protection: 12 kV impulse (1.2
µS/50 µS) based on EN 61010-1:2001 impulse

requirement for at CAT IV 1000 V/1500V dc

product. This product should not be used in

installations where transients exceed 12 kV.

AC Volts True rms
Ranges: 1000 mV, 10 V, 100 V, 1000 V
Resolution: 0.1 mV in 1000 mV range
Accuracy:
1000 mV (45 Hz to 400 Hz)

±(1.2 % rdg + 10 dgt)
10 V, 100 V (45 Hz to 500 Hz)

±(1.2 % rdg + 10 dgt)
10 V, 100 V (500 Hz to 2 kHz)

±(2.0 % rdg + 10 dgt)
1000 V (45 Hz to 1 kHz)

±(2.0 % rdg + 10 dgt)
Input Impedance: 10 MΩ
Conversion type: True rms, ac coupled 5 % to

100 % of range
Crest factor: ≤ 3
OL Protection: 1500 V dc or 1000 V ac rms
Transient protection: 12 kV impulse (1.2
µS/50 µS) based on EN 61010-1:2001 impulse

requirement for at CAT IV 1000 V product. This

product should not be used in installations

where transients exceed 12 kV

Advertising