Amprobe HD160C Digital-Multimeter User Manual

Page 19

Advertising
background image

13

DC Current
Ranges: 100 µA, 1000 µA, 10mA, 100mA,

400mA, 2A (Auto/Manual ranging)
Resolution: 0.01µA in 100 µA range
Accuracy:
100 µA range; ± (0.5% rdg + 10 dgts)
1000 µA to 400 mA ranges: ± (0.5% rdg + 5

dgts)
2 A ±(1.5% rdg + 10 dgts)range
Input protection: 2 A/1500 V fast blow ceramic

fuse 8×65 mm on A input , FP700
Burden voltage: µA range of 1 mV/1 µA, mA

range of 10 mV/1 mA, 2A range of 500 mV/1A,

500 mV max. (2 V max. on 1000 µA, 100 mA,

400 mA, 2 A ranges)
AC Current True rms
Ranges: 100 µA, 1000 µA, 10 mA, 100 mA,

400 mA, 2 A
Resolution: 0.01 µA in 100 µA range
Accuracy( 45 Hz to 1kHz):
100 µA to 100 mA: ± (1.5 % + 10 dgts)
400 mA: ± (2.0 % + 10 dgts)
2 A: ± (2.5 % + 20 dgts)
Voltage burden: see DC Current
Conversion type: True rms ac coupled 10 to

100% of range
Crest factor: ≤ 3
OL protection: see DC Current.
Resistance
Ranges:1000 Ω, 10 kΩ, 100 kΩ, 1000 kΩ,

10 MΩ, 40 MΩ
Resolution: 0.1 Ω in 1000 Ω range
Accuracy:

1000 Ω to 1000 KΩ ranges: ±(0.5 % rdg

+8 dgts);
10 MΩ range: ± (1.0 % rdg +10 dgts)
40 MΩ range: ± (2.0 % rdg +10 dgts)
Overload protection, all ranges: 1500 V dc

or 1000 V ac rms
Continuity Test
Audible indication: Less than 40 Ω
Response time: 100 ms
Overload protection:1500 V dc or 1000 V ac rms
Capacitance
Ranges: 40 ηF, 400 ηF, 4 µF, 40 µF, 400 uF (3999

counts) ( Auto/Manual ranging )
Resolution: 0.01 ηF

Accuracy: ±(3.0% rdg +10dgts) on 40 ηF, 400 µF

ranges
±(3.0% rdg +5dgts) on 400 ηF to 40 µF ranges
Test voltange: < 1 V
Test Frequency: 1.3 Hz on 40 ηF to 40 µF ranges;

0.7 Hz on 400 µF range
Overload protection: 1500 V dc or 1000 V ac rms
Temperature
Ranges: -20 °C ~ 1300 °C (-4° F ~ 2372 °F) 3999

counts
Resolution: 1 °C, 1 °F
Accuracy: ±(2.0% rdg +4 °C) -20 °C ~ 10 °C
±(1.0% rdg +3 °C)10 °C ~ 200 °C
±(2.0% rdg + 2 °C) 200 °C ~ 1300 °C
±(2.0% rdg + 8 °F) -4 °F ~ 50 °F
±(1.0% rdg + 6 °F) 50° F ~ 400 °F
±(2.0% rdg +4 °F)400 °F ~2372 °F
Overload protection: 1500 V dc or 1000 V ac rms
Frequency
Ranges: 100Hz, 1000Hz, 10kHz, 100kHz,

1000kHz, 10MHz
Resolution: 0.01 Hz on 100 Hz range
Accuracy: ±(0.1% rdg + 5 dgts)
Sensitivity:

3 Hz to 1 MHz: >2.5 V ac rms;
1MHz to 10MHz: >2.5V ac rms, <5V ac rms
Minimum input range:
100 Hz range >3 Hz;
1000 Hz range >30 Hz
Minimum pulse width: > 25 ns
Duty cycle limits: > 30 % and < 70 %
Overload protection: 1500 V dc or 1000 V ac rms

Optional Accessories
TL1500 Test Leads with Alligator Clips
CT235A 1000 A ac/dc Clamp
CT237A 200 A ac/dc Current Clamp
CT238A 20 A ac/dc Current Clamp
VC221B Padded Vinyl Case. Fits meter & holster.
DC205C Deluxe Hard-Shell Carry Case
DC207C Large Deluxe Hard-Shell Carry Case

with extra space for accessories
HV231-10A High Voltage Probe
FP 700 Replacement Fuse, 2 A/1500 V
TA-1A K-type thermocouple, temperature

adapter.