Amprobe SOLAR-600 Solar-Analyzer User Manual

Page 24

Advertising
background image

17

(1) Time delay before scan.

This delay allows light source to illuminate the solar panel

before scan stars.

(2) Sampling time of data logging (0 to 99 minutes)

(3) Current Range of Scan begin.

The beginning value of current for scan to start.

(4) Current Range of Scan end.

The ending value of current for scan to stop.

(5) Area of Solar Cell or Panel.

Based upon the input area and irradiance, this unit can

calculate the solar power converting efficiency

(6) Irradiance.

Light intensity in W/m2.

(7) Single Test Point.

Users can enter a specific value of current here. When users

press the TEST button, this specific value of current will be

simulated and result will be shown.

(8) Alarm of Low Power.

If the maximum power is lower than this value, then a beep sound will
be heard.

3. Press

or

buttons to select the setting items.

Press

or

buttons to change the setting values, or press them for

few seconds to promptly change setting values.

4. After setting Parameters, press

button to exit SETUP menu.